Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry
Purpose
To study the agreement between ultrasound (US) and Orbscan II (Bausch & Lomb) measurements with different ranges of central thickness in normal corneas.
Setting
Noor Ophthalmology Research Center, Tehran, Iran.
Methods
The central corneal thickness was measured first by the Orbscan II and then with a US pachymeter in 177 right eyes. Data were compared in 3 thickness groups: less than 500 μm, 500 to 600 μm, and more than 600 μm. In all cases, uncorrected Orbscan II data were used. For each range, a correction factor was calculated through regression analysis, which was then used in a new set of analyses. To study the agreement between the 2 methods, the 95% limits of agreement (LoA) and intraclass correlations coefficients (ICC) were determined.
Results
The mean interdevice difference was 2.7 ± 16.9 (SD) in the less than 500 μm group (P = .365), 21.3 ± 3.1 in the 500 to 600 μm group (P = .0001), and 27.2 ± 20.9 in the more than 600 μm group (P = .0001). No clinically acceptable LoA were found in any group. The application of regression equations to the Orbscan II readings improved the agreement in the less than 500 μm group better than the other 2 groups. The ICC was 0.76, 0.61, and 0.43 in the less than 500 μm, 500 to 600 μm, and more than 600 μm groups, respectively.
Conclusions
The Orbscan corneal thickness readings had relatively lower validity than US measurements. The 2 devices should not be used interchangeably, especially when the validity of the measurement is vital. In patient screening for corneal surgery, it is advisable to recheck corneal thickness with a US pachymeter.
To access this article, please choose from the options below
No author has a financial or proprietary interest in any material or method mentioned.
PII: S0886-3350(07)01171-6
doi:10.1016/j.jcrs.2007.05.036
© 2007 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Volume 33, Issue 10 , Pages 1694-1700, October 2007