Journal of Cataract & Refractive Surgery
Volume 30, Issue 6 , Pages 1320-1325, June 2004

Retrospective analysis of actual LASIK flap diameter compared with microkeratome ring size performed by different surgeons☆☆

  • Ori Mahler, MD

      Affiliations

    • Corresponding Author InformationReprint requests to Ori Mahler, MD, or Howard V. Gimbel, MD, Gimbel Eye Centre, 450, 4935-40 Avenue, NW, Calgary, Alberta T3A 2N1, Canada.
    • Gimbel Eye Centre, Calgary, Alberta, Canada
  • ,
  • Sandra J Sofinski, MD,JD

      Affiliations

    • Gimbel Eye Centre, Calgary, Alberta, Canada
  • ,
  • Howard V Gimbel, MD,MPH

      Affiliations

    • Gimbel Eye Centre, Calgary, Alberta, Canada
    • Department of Ophthalmology, Loma Linda University, Loma Linda, CA, USA
  • ,
  • Jacinthe Kassab, MD

      Affiliations

    • Gimbel Eye Centre, Calgary, Alberta, Canada
  • ,
  • Ellen E.Anderson Penno, MD

      Affiliations

    • Gimbel Eye Centre, Calgary, Alberta, Canada
  • ,
  • John A van Westenbrugge, MD

      Affiliations

    • Gimbel Eye Centre, Calgary, Alberta, Canada

Accepted 15 October 2003.

Abstract 

Purpose: To evaluate the difference in the diameter between microkeratome ring and actual laser in situ keratomileusis (LASIK) flap size by 4 different surgeons using 3 different microkeratomes in comparison to preoperative keratometry.

Setting: Gimbel Eye Centre, Calgary, Alberta, Canada.

Methods: A retrospective analysis of 6667 consecutive LASIK surgeries performed at the Gimbel Eye Centre between 1998 and 2002 was conducted. The study compared the difference in the actual LASIK flap diameter and microkeratome ring size for 3 Hansatome microkeratomes indicated by serial numbers 2230, 3048, and 3084 in comparison to the preoperative keratometry. The surgeries were performed by 4 surgeons from the Centre.

Results: A statistically significant correlation was found between the actual flap size and the microkeratome ring size. The deviation increased with corneal steepness (#2230: r = 0.45, P<.001; #3048: r = 0.28, P<.001; #3084: r = 0.23, P<.001). The mean difference between actual flap size and the microkeratome ring size was 0.647 mm, 0.808 mm, and 0.916 mm for Hansatome microkeratome #2230, #3048, and #3084, respectively. The actual flap was larger than the microkeratome ring size in 92.8%, 93.8%, and 97.4% of eyes, respectively. The mean difference between microkeratome ring size and measured LASIK flap size was consistently significantly different (P<.0001) per surgeon for all microkeratomes.

Conclusions: Estimation of the flap size is an important issue while selecting ring size and planning the treatment zone. In this study, the actual flap size was larger than the ring size for all surgeons and all 3 Hansatome microkeratomes. Higher preoperative keratometry values achieved larger flaps than those with lower keratometry values.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Presented in part at the ASCRS Symposium on Cataract, Refractive and IOL Surgery, April 2003, San Francisco, California, USA.

☆☆ The authors have no propriety or financial interest in the medical products or devices mentioned.

PII: S0886-3350(03)01105-2

doi:10.1016/j.jcrs.2003.10.030

Journal of Cataract & Refractive Surgery
Volume 30, Issue 6 , Pages 1320-1325, June 2004