Journal of Cataract & Refractive Surgery
Volume 30, Issue 9 , Pages 1993-1997, September 2004

Spring constants of capsular tension rings

  • Sabine Kurz, MD

      Affiliations

    • Corresponding Author InformationReprint requests to Dr. med. Sabine Kurz, Department of Ophthalmology, Johannes Gutenberg-University, Langenbeckstrasse 1, 55101 Mainz, Germany.
  • ,
  • H. Burkhard Dick, MD

From the Department of Ophthalmology, Johannes Gutenberg-University, Mainz, Germany

Accepted 14 January 2004.

Purpose

To determine the spring constants of several types of capsular tension rings (CTRs).

Setting

Department of Ophthalmology, Johannes Gutenberg-University, Mainz, Germany.

Methods

Fourteen models of CTRs provided by 9 companies were tested; all were 11.0 mm in diameter. The CTRs were positioned in a special retainer of a material-testing machine and compressed at a constant speed of 10.0 mm per minute. The distance and force values were recorded, and each CTR model was tested 3 times using a new specimen each time. The resulting measurements were used to calculate a mean spring constant for each CTR model.

Results

All tested CTRs showed linear proportionality between force and compression distance. Test results were repeatable and precise. Capsular tension rings of the same model had comparable mechanical properties, and different models had different mean spring constants ranging from 0.8162 to 4.5501 mN/mm.

Conclusions

The spring constant of a CTR is a suitable mechanical characteristic to facilitate the choice of CTR model. A CTR with a low spring constant may be advantageous for the management of zonular dialysis, whereas a CTR with a higher spring constant may reduce capsular bag shrinkage more effectively.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 None of the authors has a financial or proprietary interest in any material or method mentioned.

PII: S0886-3350(03)01155-6

doi:10.1016/j.jcrs.2003.12.031

Journal of Cataract & Refractive Surgery
Volume 30, Issue 9 , Pages 1993-1997, September 2004