Journal of Cataract & Refractive Surgery
Volume 33, Issue 2 , Pages 210-216, February 2007

Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation:

Effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision

From the Department of Ophthalmology, University Hospital of Tours, Tours, France

Accepted 17 October 2006.

Purpose

To compare ocular performance and quality of vision in pseudophakic eyes with an aspherical intraocular lens (IOL) or a conventional spherical IOL.

Setting

Bretonneau University Hospital, Tours, France.

Methods

Twenty patients (40 eyes) were randomly divided in 2 equal groups to bilaterally receive the aspherical Tecnis Z9000 IOL (AMO) or the spherical CeeOn Edge 911 IOL (AMO). Contrast sensitivity was measured and ocular wavefront analysis performed before surgery and 6 months after. Patients completed the Activities of Daily Vision Scale (ADVS) to evaluate patient-centered visual outcomes. Other examinations included refraction before and after mydriasis and pupil diameter.

Results

The mean postoperative best corrected visual acuity (logMAR) was 0.03 ± 0.05 (SD) in the Tecnis group and 0.01 ± 0.05 in the CeeOn Edge group (P = .41). Refractive evaluation with mydriasis showed a mean myopic shift as low as −0.02 ± 0.36 diopter (D) in the Tecnis group and −0.51 ± 0.37 D in the CeeOn Edge group (P = .001). Mesopic contrast sensitivity at high spatial frequencies was significantly better in the Tecnis group (P<.001), while contrast sensitivity under photopic and glare conditions was not different between the 2 groups. Spherical aberration was significantly lower in the Tecnis group, which had a mean Z40 of 0.01 ± 0.06 μm, than in the CeeOn Edge group, which had a mean Z40 of 0.16 ± 0.12 μm (P<.001). The global score on the ADVS was not statistically different between groups; however, quality of distance vision was better in the Tecnis group than in the CeeOn Edge group (mean 99.0 ± 2.0 versus 89.2 ± 3.4) (P<.001).

Conclusion

Implantation of an aspherical IOL with a negative spherical aberration resulted in reduced ocular spherical aberration and improved mesopic contrast sensitivity and led to better subjective quality of vision.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 No author has a financial or proprietary interest in any material or method mentioned.

PII: S0886-3350(06)01421-0

doi:10.1016/j.jcrs.2006.10.035

Journal of Cataract & Refractive Surgery
Volume 33, Issue 2 , Pages 210-216, February 2007