Journal of Cataract & Refractive Surgery
Volume 33, Issue 6 , Pages 1041-1044, June 2007

Scheimpflug imaging to determine intraocular lens power in vivo

From the King's College Hospital Ophthalmology Department, King's College Hospital, London, United Kingdom

Accepted 19 February 2007.

Purpose

To assess the feasibility of determining intraocular lens (IOL) power by measurement of the central optic thickness using clinically available Scheimpflug imaging (Pentacam, Oculus).

Setting

King's College Hospital Ophthalmology Department, London, United Kingdom.

Methods

Sixty-seven eyes were assessed 1 month after uneventful phacoemulsification with in-the-bag implantation of AcrySof MA60AC IOLs (Alcon). The correlation between IOL thickness measurement and IOL power was calculated. Repeatability of central optic thickness measurement was determined from 10 successive scans of 4 patients.

Results

Within-subject standard deviation increased with the subject mean. The coefficient of variability was 1.4%. Measured lens thickness was highly correlated with lens power (R2 = 0.94, P<.001). Over the measured range, 95% confidence intervals varied between ±0.83 diopters (D) and ±0.92 D.

Conclusions

Central IOL thickness measurements with the Pentacam Scheimpflug camera were highly repeatable and closely correlated with the known IOL power. The IOL power, calculated from a regression equation, is likely to be less than ±1.00 D away from the actual power. Approximate in vivo IOL power determination is feasible with clinically available Scheimpflug imaging. This could be applied clinically in cases of unexplained postoperative refractive error.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 No author has a financial or proprietary interest in any material or method mentioned.Presented at the Imaging in the Eye III: Technologies and Clinical Applications meeting at the Institute of Physics, London, England, September 2006.Funded by the Ophthalmology Department, King's College Hospital, London, United Kingdom.Temi Adewoyin, MRCOphth, and Mike R.F. Aitken, PhD, provided statistical assistance.

PII: S0886-3350(07)00449-X

doi:10.1016/j.jcrs.2007.02.023

Journal of Cataract & Refractive Surgery
Volume 33, Issue 6 , Pages 1041-1044, June 2007