Journal of Cataract & Refractive Surgery
Volume 34, Issue 5 , Pages 763-768, May 2008

Optical quality of the eye after lens replacement with a pseudoaccommodating intraocular lens

From the Optometry Research Group (Montés-Micó, Ferrer-Blasco, Cerviño), Optics Department, University of Valencia, and the Fernández-Vega Ophthalmological Institute (Alfonso, Fernández-Vega), Valencia, and the Surgery Department (Alfonso, Fernández-Vega), School of Medicine, University of Oviedo, Oviedo, Spain; Faculty of Life Sciences (Charman), University of Manchester, Manchester, United Kingdom

Accepted 8 January 2008.

Purpose

To examine the optical quality of healthy presbyopic eyes before and after implantation of the pseudoaccommodating AcrySof ReSTOR Natural intraocular lens (IOL) (Alcon Laboratories) by measuring visual acuity and wavefront aberrations.

Setting

University of Valencia, Fernández-Vega Ophthalmological Institute, Valencia, Spain.

Methods

Monocular refraction and photopic distance (6 m) and near (0.33 m) visual acuities were determined before and after clear lens extraction with bilateral AcrySof ReSTOR Natural IOL implantation in 30 hyperopic eyes (+0.50 to +4.00 diopters [D]) of presbyopic patients with low astigmatism (≤1.00 D). Postsurgical binocular visual acuity was measured over a range of effective distances. Wavefront aberrations were measured before and after surgery. Aberrations for 2 pupil diameters (3.0 mm and 5.0 mm) were decomposed using Zernike polynomials; total, defocus, astigmatism spherical, and coma aberration terms were studied separately. Postimplantation results were compared with preoperative values.

Results

Safety and efficacy indices for distance vision were 1.01 and 0.96, respectively. At near, they improved to 1.07 and 1.06, respectively. Vision at intermediate distances was noticeably poorer. Total, defocus, and spherical aberrations changed after IOL implantation with both pupil diameters (P<.01). Total and defocus aberrations were significantly reduced postoperatively. Spherical aberration coefficients changed to negative and less positive values for a 3.0 mm and a 5.0 mm pupil, respectively.

Conclusions

Implantation of the pseudoaccommodating IOL in healthy eyes yielded uncorrected levels of high-contrast, distance, and near acuities comparable to preoperative best corrected values. Higher-order aberration levels appeared to be similar to the preoperative values.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 No author has a financial or proprietary interest in any material or method mentioned.

 Supported in part by a Universitat de Valencia research grant (UV-AE-20070225) (Dr. Montés-Micó).

PII: S0886-3350(08)00194-6

doi:10.1016/j.jcrs.2008.01.017

Journal of Cataract & Refractive Surgery
Volume 34, Issue 5 , Pages 763-768, May 2008