In the paper entitled “New Intraocular Lens for Achromatizing the Human Eye” (J Cataract Refract Surg 2007; 33:1296–1302), there are 2 small errors that may affect the outcomes if one tries to perform some calculations. On page 1298, the equation in the last paragraph should not have parentheses; that is,
. On page 1299, the data for “Aqueous” and “Intraocular Lens” in Table 2 are incorrect. The correct table is as follows:
Table 2. Values of no, A, and B used in the model to modulate the refraction index variation with wavelength.
| Value | Cornea | Aqueous | Intraocular Lens | Vitreous |
|---|---|---|---|---|
| no | 1.36468149 | 1.32628942 | 1.4732734 | 1.32468149 |
| A | 0.00478075 | 0.00441874 | 0.00918165 | 0.00478075 |
| B | 0.00049921 | 0.00055639 | 0.000447667 | 0.00049921 |
PII: S0886-3350(08)00434-3
doi:10.1016/j.jcrs.2008.05.007
© 2008 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Refers to article:
- New intraocular lens for achromatizing the human eye