Apodized diffractive versus refractive multifocal intraocular lenses: Optical and visual evaluation

Beata Żelichowska, MD, Marek Rękas, PhDCorresponding Author Informationemail address, Andrzej Stankiewicz, PhD, Alejandro Cerviño, PhD, Robert Montés-Micó, PhD

Accepted 27 June 2008.

Purpose

To evaluate the optical and visual performance after implantation of refractive or apodized diffractive multifocal intraocular lenses (IOLs).

Setting

Military Health Service Institute, Warsaw, Poland.

Methods

Uncorrected distance visual acuity, best distance-corrected visual acuity, best distance-corrected near visual acuity, distance contrast sensitivity under photopic conditions (CSV-1000), residual refractive error, and wavefront aberrations (LADARWave Hartmann-Shack wavefront analyzer) were measured in 23 patients who had bilateral implantation of the AcrySof ReSTOR SN60D3 IOL and 23 patients who had bilateral implantation of the ReZoom IOL….

Results

At the 6-month postoperative visit, the mean photopic uncorrected distance acuity was 0.03 ± 0.05 (SD) in the ReSTOR group and 0.02 ± 0.06 logMAR in the ReZoom group (both approximately 20/20) (P = .569). In all patients, the mean photopic best distance-corrected acuity was 0.00 logMAR (approximately 20/20) and the mean photopic best distance-corrected near acuity at 35 cm was 0.10 logMAR. The photopic contrast sensitivity was within the standard normal range in both IOL groups. The difference in photopic contrast sensitivity between groups was statistically significant (P<.001). Higher-order aberrations, in particular coma and spherical aberrations, were significantly higher in the ReZoom group (all P<.001).

Conclusions

AcrySof ReSTOR SN60D3 and ReZoom IOLs provided good visual performance at distance and near under photopic conditions. Optical quality measures were significantly worse in patients with ReZoom IOLs.

From the Ophthalmology Department (Żelichowska, Rękas, Stankiewicz), Military Health Service Institute, Warsaw, Poland; Optometry Research Group (Cerviño, Montés-Micó), Optics Department, University of Valencia, Valencia, Spain