Influence of anterior chamber morphometrics on endothelial cell changes after phakic intraocular lens implantation

Muriël Doors, MD, Diana W.J.K. Cals, MD, Tos T.J.M. Berendschot, PhD, John de Brabander, PhD, Fred Hendrikse, MD, PhD, Carroll A.B. Webers, MD, PhD, Rudy M.M.A. Nuijts, MD, PhDCorresponding Author Informationemail address

Accepted 20 August 2008.

Purpose

To analyze the position of iris-fixated phakic intraocular lenses (pIOLs) using anterior segment optical coherence tomography (AS-OCT) and evaluate the effect of anterior chamber morphometrics on endothelial cell changes.

Setting

Department of Ophthalmology, Academic Hospital Maastricht, The Netherlands.

Methods

In this cross-sectional study, AS-OCT was used to measure the distances from the center and the edges of the pIOL to the corneal endothelium in 242 eyes with various models of myopic pIOLs. Endothelial cell measurements were performed preoperatively and at each follow-up examination.

Results

The mean follow-up was 34.1 months ± 24.7 (SD) (range 3 months to 7 years). The mean distance between the edge of the pIOL and the endothelium was 1.37 ± 0.22 mm. Although this distance was smaller than the safety value of 1.50 mm in 68.6% of the eyes, no eye developed corneal decompensation. There was a significant endothelial cell density (ECD) loss of 1.28% ± 8.46%, 3.25% ± 8.24%, and 5.02% ± 10.40% at 2 years, 5 years, and 7 years, respectively. Linear mixed-model analysis predicted a yearly ECD loss of 0.98% for a mean edge distance of 1.37 mm, 0.15% for an edge distance of 1.59 mm (mean plus 1 SD), and 1.80% for an edge distance of 1.15 mm (mean minus 1 SD).

Conclusions

A shorter distance between the edge of the pIOL and the endothelium was significantly associated with higher ECD loss. For safety reasons, the postoperative examination should include long-term evaluation of the anterior chamber morphometrics in addition to ECD counts.

From the Department of Ophthalmology, Academic Hospital Maastricht, Maastricht, The Netherlands