Spherical aberration yielding optimum visual performance: Evaluation of intraocular lenses using adaptive optics simulation
Received 6 November 2008; received in revised form 18 January 2009; accepted 25 February 2009.
Purpose
To evaluate the influence of spherical aberration on contrast sensitivity using adaptive optics.
Setting
Vision Science and Advanced Retinal Imaging Laboratory, Department of Ophthalmology & Vision Science, University of California, Davis Medical Center, Sacramento, California, USA.
Methods
Contrast sensitivity at 8 cycles per degree was evaluated using an adaptive optics system that permitted aberrations to be measured with a Hartmann-Shack wavefront sensor and controlled by a 109 actuator continuous-surface deformable mirror that was at a plane conjugate to the observer's pupil. Vertical Gabor patches were viewed through a 6.3 mm diameter pupil conjugate aperture. Contrast sensitivity was measured with the deformable mirror set to produce 1 of 5 spherical aberration profiles (−0.2 to +0.2 μm). Contrast sensitivity over the range of spherical aberration was fitted with a polynomial function.
Results
Three subjects (age 21 to 24 years) participated. The measured total mean spherical aberration resulting from the spherical aberration profiles produced by the deformable mirror was between −0.15 μm and +0.25 μm. The peak contrast sensitivity of this function for the 3 subjects combined occurred at +0.06 μm of spherical aberration. The peak contrast sensitivity was also achieved with positive spherical aberration for each subject's data fitted individually (mean 0.09).
Conclusion
There was intersubject variability in the measurements; however, the mean visual performance was best with the introduction of a small positive spherical aberration.
From the University of California Davis Eye Center, Department of Ophthalmology & Vision Science, University of California, Davis Medical Center, Sacramento, California, USA
Corresponding author: John S. Werner, UC Davis Eye Center, Department of Ophthalmology & Vision Science, University of California–Davis Medical Center, 4860 Y Street, Suite 2400, Sacramento, California 95817, USA.